CR & DR Systems: Digital Imaging Workflow & Artifacts

Image Production — 25% > Digital Imaging Systems & PACS
Learn

Overview of Digital Imaging Systems

Digital imaging has replaced film-screen systems in virtually all modern radiology departments. The ARRT Radiography exam tests your understanding of computed radiography (CR) and digital radiography (DR) workflows, image processing, exposure indicators, artifacts, and quality control protocols.

Definition: Computed Radiography (CR) uses a photostimulable phosphor (PSP) plate inside a cassette to capture the latent image, which is later read by a laser scanner. Digital Radiography (DR) uses a flat-panel detector (FPD) permanently installed in the table or wall bucky, delivering an image directly to the workstation.

Key areas to master for the exam:

  • Step-by-step CR and DR workflows
  • Image processing and PACS/DICOM/RIS integration
  • Exposure indicators and deviation index (DI)
  • Common digital artifacts and their causes
  • Quality control responsibilities

CR & DR Imaging Workflows

Image Processing & Information Systems

After acquisition, digital images undergo electronic processing before display. The technologist interacts with multiple information systems daily.

Image Processing Steps: Raw image data is processed through histogram analysis, look-up table (LUT) application, edge enhancement, and noise reduction before display on the PACS workstation.

SystemFunction
PACS (Picture Archiving and Communication System)Stores, retrieves, and distributes digital radiographic images
DICOM (Digital Imaging and Communications in Medicine)Universal standard for medical image format and communication
RIS (Radiology Information System)Manages patient scheduling, reporting, and workflow

Exam Tip: DICOM is the universal standard that allows images from different vendors to be shared across systems. PACS is the storage and distribution platform. RIS handles the administrative/scheduling side. Know the distinction — the ARRT exam tests all three.

Exposure Indicators & Quality Control

Summary & Exam Takeaways

KEY CONCEPTS — Test-Ready Recall

  • CR workflow: PSP plate → cassette → reader → laser → PMT → digital signal → erase plate with bright light.
  • DR workflow: FPD permanently installed → image available in 5–15 seconds → no cassette or erasure step.
  • Ghost image = CR only (incomplete erasure). Lag artifact = DR only (residual signal).
  • Fuji S-number is inverse — higher S = lower exposure. All other major systems (Kodak EI, AGFA lgM, Canon REI) are direct.
  • Deviation Index (DI): Target = 0. Positive DI = overexposure; negative DI = underexposure. Set by the medical physicist.
  • Quantum noise is caused by low mAs — increase mAs to reduce noise.
  • Moiré pattern = stationary grid frequency interferes with detector sampling — use a moving Bucky grid.
  • PACS = image storage/distribution. DICOM = universal image standard. RIS = scheduling/reporting.
  • QC: Flat field and artifact evaluation are daily tasks; spatial resolution and noise testing are monthly/annual tasks performed with medical physics oversight.